8.5 Air Cooling System 
The  sensing  head  can  operate  in  ambient  temperatures  up  to  200°C 
(392°F)  with  the  air‐cooling  system.  The  air‐cooling  system  comes 
with a T‐adapter including 0.8 m / 31.5 in (optional: 2.8 m / 110 in) air 
hose and insulation. The T‐adapter allows the air‐cooling hose to be 
installed without interrupting the connections to the box. 
The air‐cooling jacket may be combined with the right angle mirror. 
Figure 35: Air Cooling System 
 
Figure 36: Connecting the T‐Adapter (XXXMIACCJ) 
max. ambient 200°C (392°F) 
Sensing Head 
max. ambient 50°C (122°F) 
Air Hose 
T-
Adapter 
Electronic 
Housing 
Cable 
Air cooling (max. 35°C / 95°F) 
Hose to 
sensing head 
T-Adapter 
Cable to electronic 
housing 
Fitting free for air connection 
Hose: 
inner Ø: 9 mm (0.35 in) 
outer Ø: 12 mm (0.47 in)