thawing time may vary. The amount of air left in the plasma 
thawing bag during thawing will also affect thawing time. 
The laboratory personnel can set the desired thawing time for 
each unit to be thawed. They do not have to all be set for the 
same thaw time. 
 
  f.  To change the default TIMER : 
 
Press the START/CHANGE key for the selected bag location, 
then press the numeric key pad for the desired minutes. All 
keys must be pressed within 5 seconds of each other. Minutes 
must be between 1 and 99. 
 
  g.  To start rocking motion of the bath, press the ROCK key. 
 
Rock indicator will light. Pressing the ROCK key at any time 
will stop the rocking motion. 
 
      NOTE : If the water spills over the top of the bath when 
rocking motion begins, stop the bath from rocking and 
remove a small amount of water. 
 
 
  h. To thaw an additional bag, place it in slot 2, 3, etc.    Adjust 
the timer as required, follow instructions in Section 2.5-f 
above. 
 
The product should be thawed when an audible signal is 
heard and the respective BAG TIMER is at zero. Remove the 
product from that slot location, check the bag contents to 
ensure that it is completely thawed. If additional thawing is 
required, return the bag to the holder and set timer for the 
desired thaw time. 
 
 
 
 
 
 