SET PUMP STAGE TIME?
       NO/YES 
 
STAGE DELAY:  3.000s
PRESS ENTER TO SAVE 
 
Setting the Pump Stage Time 
Pump stage time
 is a delay that prevents both pumps from engaging at the same time. 
Although rare in a two-pump system, both pumps can sometimes start at the same time. This 
condition causes an excess of current on the electrical system from such a surge. 
The pump stage time delay defaults to 3.0 seconds and should not be altered unless you are 
instructed to do so by the Service Dept. 
 
 
 
 
 
 
 
At the above message prompt: 
1.  Press the YES key to enter the Pump Stage Time menu. 
 
 
 
 
 
 
 
 
Using the keypad: 
2.  Enter the new time and press the ENTER key to save the new setting.