Teledyne 200A User Manual

Page of 225
Teledyne API Model 200A NO
X
 Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 
 
8
  
MAINTENANCE 
8.1  Maintenance Schedule 
NOTE 
The operations outlined in this chapter are to be  
performed by qualified maintenance personnel only. 
 
Table 8-8-1:  Preventative Maintenance Schedule 
Item 
Maintenance Interval 
Reference Section 
TEST Functions 
6 - 12 month intervals 
Zero/Span Calibration 
Annually or after repairs 
Section 
Zero/Span Checks 
Daily 
Particulate Filter 
Weekly as needed 
Ozone Scrubber on Pump Pack 
Replace charcoal every 3 months Section 4.2.7Table 4-4-3 
Filter for Ozone Scrubber 
Replace every 12 months 
Filter for PermaPure Drier 
Replace every 12 months 
Zero Air Canister 
Refill every 3 months 
Section 8.5 
Reaction Cell Window 
Clean annually or as necessary 
Section 9.3.8Table 8-8-5 
Ozone Flow 
Check every year 
Sample Flow 
Check every year 
Moly Converter 
Check efficiency every 6 months Table 8-8-4, Section 8.6 
Pneumatic Lines 
Examine every 12 months, clean 
if necessary 
Factory Calibration  
Calibrate each year or after 
repairs 
Section 9.1.6 
Leak Check 
Check every year 
Section 8.9 
O-rings 
Replace if seal is broken 
8-1 