Teledyne 200A User Manual
Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
8
MAINTENANCE
8.1 Maintenance Schedule
NOTE
The operations outlined in this chapter are to be
performed by qualified maintenance personnel only.
Table 8-8-1: Preventative Maintenance Schedule
Item
Maintenance Interval
Reference Section
TEST Functions
6 - 12 month intervals
Zero/Span Calibration
Annually or after repairs
Section 7
Zero/Span Checks
Daily
Section 7, Table 6-6-2
Particulate Filter
Weekly as needed
Ozone Scrubber on Pump Pack
Replace charcoal every 3 months Section 4.2.7, Table 4-4-3
Filter for Ozone Scrubber
Replace every 12 months
Filter for PermaPure Drier
Replace every 12 months
Zero Air Canister
Refill every 3 months
Section 8.5
Reaction Cell Window
Clean annually or as necessary
Section 9.3.8, Table 8-8-5
Ozone Flow
Check every year
Sample Flow
Check every year
Moly Converter
Check efficiency every 6 months Table 8-8-4, Section 8.6
Pneumatic Lines
Examine every 12 months, clean
if necessary
Factory Calibration
Calibrate each year or after
repairs
Section 9.1.6
Leak Check
Check every year
Section 8.9
O-rings
Replace if seal is broken
8-1