Agilent Technologies E7405A User Manual

Page of 92
12
Chapter 1
 
Making Basic Measurements
Comparing Signals
Figure 1-2
Using the Marker Delta Function 
Signal Comparison Example 2:
Measure the frequency and amplitude difference between two signals 
that do not appear on the screen at one time. (This technique is useful 
for harmonic distortion tests when narrow span and narrow bandwidth 
are necessary to measure the low level harmonics.)
1.
Perform a factory preset by pressing 
Preset
Factory Preset
 (if 
present). 
2.
Connect the 10 MHz REF OUT from the rear panel to the 
front-panel INPUT.
3.
Set the center frequency to 10 MHz by pressing 
FREQUENCY
Center Freq
, 10, 
MHz
4.
Set the span to 5 MHz by pressing 
SPAN,
 5, 
MHz
5.
Set the resolution bandwidth to spectrum analyzer coupling by 
pressing 
BW/Avg
Res BW
 (SA). 
6.
Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
7.
Set the reference level to 10 dBm by pressing 
AMPLITUDE
Ref Level
10, 
dBm
The 10 MHz reference signal appears on the display. 
8.
Press 
Peak Search
 to place a marker on the peak.
9.
Press 
Marker
Mkr
CF Step
 to set the center frequency step size 
equal to the frequency of the fundamental signal.