Agilent Technologies E7405A User Manual

Page of 92
Chapter 1
15
Making Basic Measurements
Resolving Signals of Equal Amplitude
Resolving Signals Example:
Resolve two signals of equal amplitude with a frequency separation of 
100 kHz. 
1.
Connect two sources to the analyzer input as shown in 
.
Figure 1-4
Setup for Obtaining Two Signals
2.
Set one source to 300 MHz. Set the frequency of the other source to 
300.1 MHz. The amplitude of both signals should be approximately 
20 dBm at the output of the bridge. 
3.
Set the analyzer as follows:
a.
Press 
Preset
Factory Preset
 (if present).
b.
Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
c.
Set the center frequency to 300 MHz by pressing 
FREQUENCY,
 
Center Freq
, 300, 
MHz
.
d.
Set the span to 2 MHz by pressing 
SPAN,
 
Span
, 2, 
MHz
.
e.
Set the resolution bandwidth to 300 kHz by pressing 
BW/Avg
Res BW
, 300, 
kHz
A single signal peak is visible. See 
NOTE
If the signal peak is not present on the display, do the following:
1.
Increase the span to 20 MHz by pressing 
SPAN,
 
Span
, 20, 
MHz
The signal should be visible. 
2.
Press 
Peak Search
FREQUENCY
Signal Track
 (On)
3.
Press 
SPAN
, 2, 
MHz
 to bring the signal to center screen. 
4.
Press 
FREQUENCY
Signal Track
 (Off)