Chapter 1
 
Making Basic Measurements
Decreasing the Frequency Span Around the Signal
Decreasing the Frequency Span Around the 
Signal 
Using the analyzer signal track function, you can quickly decrease the 
span while keeping the signal at center frequency. This is a fast way to 
take a closer look at the area around the signal to identify signals that 
would otherwise not be resolved.
Decreasing the Frequency Span Example:
Examine a signal in a 200 kHz span. 
1.
Perform a factory preset by pressing 
Preset
Factory Preset
 (if 
present). 
2.
Turn on the internal 50 MHz amplitude reference signal of the 
analyzer as follows: 
For the E7401A, use the internal 50 MHz amplitude reference 
signal of the analyzer as the signal being measured. Press 
Input/Output
Amptd Ref
 (On). 
For all other models connect a cable between the front-panel 
AMPTD REF OUT to the analyzer INPUT, then press 
Input/Output
Amptd Ref Out
 (On).
3.
Set the start frequency to 20 MHz by pressing 
FREQUENCY
Start Freq
, 20, 
MHz
.
4.
Set the stop frequency to 1 GHz by pressing 
FREQUENCY
Stop Freq
1, 
GHz
.
5.
Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
6.
Set the resolution bandwidth to spectrum analyzer coupling pressing 
BW/Avg
Resolution BW
 (SA). 
7.
Press 
Peak Search
 to place a marker at the peak. See