Chapter 1
27
Making Basic Measurements
Tracking Drifting Signals
Tracking Drifting Signals 
The signal track function is useful for tracking drifting signals that 
drift relatively slowly. To place a marker on the signal you wish to 
track, use 
Peak Search
. Pressing 
FREQUENCY
Signal Track
 (On) will 
bring that signal to the center frequency of the graticule and adjust the 
center frequency every sweep to bring the selected signal back to the 
center. A quick way to perform the 
Peak Search
FREQUENCY
Signal Track
SPAN
 key sequence is to use the 
Span Zoom
 key in the 
SPAN
 menu.
Note that the primary function of the signal track function is to track 
unstable signals, not to track a signal as the center frequency of the 
analyzer is changed. If you choose to use the signal track function when 
changing center frequency, check to ensure that the signal found by the 
tracking function is the correct signal. 
Tracking Signal Drift Example 1:
Use the signal track function to keep a drifting signal at the center of 
the display and monitor its change. 
This example requires a signal generator. The frequency of the signal 
generator will be changed while you view the signal on the display of 
the analyzer. 
1.
Connect a signal generator to the analyzer input. 
2.
Set the signal generator frequency to 300 MHz with an amplitude of 
20 dBm. 
3.
Set the analyzer as follows:
a.
Press 
Preset
Factory Preset
 (if present).
b.
Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
c.
Set the resolution bandwidth to the spectrum analyzer coupling 
by pressing 
BW/Avg
Resolution BW
 (SA). See 
d.
Set the center frequency to 300 MHz by pressing 
FREQUENCY,
 
Center Freq
, 300, 
MHz
.