Agilent Technologies E7405A User Manual

Page of 92
30
Chapter 1
 
Making Basic Measurements
Tracking Drifting Signals
8.
The signal frequency drift can be read from the screen if both the 
signal track and marker delta functions are active. Set the analyzer 
and signal generator as follows:
a.
Press 
Marker, Delta
b.
Tune the frequency of the signal generator. The marker readout 
indicates the change in frequency and amplitude as the signal 
drifts. See 
Figure 1-21
Using Signal Tracking to Track a Drifting Signal 
Tracking Signal Drift Example 2:
The analyzer can measure the short- and long-term stability of a 
source. The maximum amplitude level and the frequency drift of an 
input signal trace can be displayed and held by using the 
maximum-hold function. You can also use the maximum hold function if 
you want to determine how much of the frequency spectrum a signal 
occupies. 
1.
Connect a signal generator to the analyzer input. 
2.
Set the signal generator frequency to 300 MHz with an amplitude of 
20 dBm. 
3.
Set the analyzer as follows:
a.
Press 
Preset
Factory Preset
 (if present).
b.
Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
c.
Set the resolution bandwidth to spectrum analyzer coupling by 
pressing 
BW/Avg
Resolution BW
 (SA).