Agilent Technologies E7405A User Manual

Page of 92
34
Chapter 1
 
Making Basic Measurements
Measuring Low Level Signals
Measuring Low Level Signals
The ability of the analyzer to measure low level signals is limited by the 
noise generated inside the analyzer. A signal may be masked by the 
noise floor so that it is not visible. This sensitivity to low level signals is 
affected by the measurement setup.
The analyzer input attenuator and bandwidth settings affect the 
sensitivity by changing the signal-to-noise ratio. The attenuator affects 
the level of a signal passing through the instrument, whereas the 
bandwidth affects the level of internal noise without affecting the 
signal. In the first two examples in this section, the attenuator and 
bandwidth settings are adjusted to view low level signals. 
If, after adjusting the attenuation and resolution bandwidth, a signal is 
still near the noise, visibility can be improved by using the video 
bandwidth and video averaging functions, as demonstrated in the third 
and fourth examples. 
Measuring Low Level Signals Example 1:
If a signal is very close to the noise floor, reducing input attenuation 
brings the signal out of the noise. Reducing the attenuation to 0 dB 
maximizes signal power in the analyzer. 
CAUTION
The total power of all input signals at the analyzer input must not 
exceed the maximum power level for the analyzer. 
1.
Connect a signal generator to the analyzer input. 
2.
Set the signal generator frequency to 300 MHz with an amplitude of 
80 dBm. 
3.
On the analyze, perform a factory preset by pressing 
Preset
Factory Preset
 (if present). 
4.
Set the center frequency of the analyzer to 300 MHz by pressing 
FREQUENCY
Center Freq
, 300, 
MHz
5.
Set the span to 5 MHz by pressing 
SPAN
Span
, 5, 
MHz
6.
Set the resolution bandwidth to spectrum analyzer coupling by 
pressing 
BW/Avg
Res BW
 (SA). 
7.
Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
8.
Set the reference level to 
40 dBm by pressing 
AMPLITUDE
Ref Level
–40, 
dBm
.
9.
Place the signal at center frequency by pressing 
Peak Search
Marker
Mkr
CF