Chapter 1
 
Making Basic Measurements
Identifying Distortion Products
Identifying Distortion Products 
Distortion from the Analyzer 
High level input signals may cause analyzer distortion products that 
could mask the real distortion measured on the input signal. Using 
trace 2 and the RF attenuator, you can determine which signals, if any, 
are internally generated distortion products.
Identifying Analyzer Generated Distortion Example:
Using a signal from a signal generator, determine whether the 
harmonic distortion products are generated by the analyzer. 
1.
Connect a signal generator to the analyzer INPUT. 
2.
Set the signal generator frequency to 200 MHz and the amplitude to 
0 dBm. 
3.
On the analyzer, perform a factory preset by pressing 
Preset
Factory Preset
 (if present).
4.
Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
5.
Set the resolution bandwidth to spectrum analyzer coupling by 
pressing 
BW/Avg
Res BW
 (SA). 
6.
Set the center frequency of the analyzer to 400 MHz by pressing 
FREQUENCY
Center Freq
, 400, 
MHz
7.
Set the span to 500 MHz by pressing 
SPAN
Span
, 500, 
MHz
The signal produces harmonic distortion products in the analyzer 
input mixer as shown in