Agilent Technologies E7405A User Manual

Page of 92
56
Chapter 1
 
Making Basic Measurements
Making Noise Measurements
Noise Measurement Example 2:
The Normal marker can also be used to make a single frequency 
measurement as described in the previous example, again using video 
filtering or averaging to obtain a reasonably stable measurement. 
While video averaging automatically selects the sample display 
detection mode, video filtering does not. With sufficient filtering that 
results in a smooth trace, there is no difference between the sample and 
peak modes because the filtering takes place before the signal is 
digitized.
Be sure to account for the fact that the averaged noise is displayed 
approximately 2 dB too low for a noise bandwidth equal to the 
resolution bandwidth. Therefore, you must add 2 dB to the marker 
reading. For example, if the marker indicates –100 dBm, the actual 
noise level is –98 dBm.
Noise Measurement Example 3:
You may use adjustable markers to set the frequency span over which 
power is measured. The markers allow you to easily and conveniently 
select any arbitrary portion of the displayed signal for measurement. 
However, while the analyzer does select the average (video/rms) display 
detection mode, you must set all of the other parameters. 
1.
Reset the analyzer by pressing 
Preset
Factory Preset
 (if present).
2.
Tune the analyzer to the frequency of 50 MHz. In this example we 
are using the amplitude reference signal. Press 
FREQUENCY
, 50, 
MHz
.
3.
Set the span to 100 kHz by pressing 
SPAN
, 100, 
kHz
.
4.
Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
5.
Set the resolution bandwidth to 1 kHz by pressing 
BW/Avg
Resolution BW
, 1, 
kHz
.
6.
Set the reference level to 
20 dBm by pressing 
AMPLITUDE
Ref Level
–20, 
dBm
7.
Set the input attenuator to 40 dB by pressing 
Attenuation
, 40, 
dB
.
8.
Set the marker span to 40 kHz by pressing 
Marker
Span Pair
 (Span), 
40, 
kHz
.
The resolution bandwidth should be about 1 to 3% of the 
measurement (marker) span, 40 kHz in this example. The 1 kHz 
resolution bandwidth that the analyzer has chosen is fine. The video 
bandwidth should be ten times wider.
9.
Set the video bandwidth to 10 kHz by pressing 
BW/Avg
Video BW
(Man), 10, 
kHz
.