Making Basic Measurements
Demodulating AM Signals (Using the Analyzer As a Fixed Tuned Receiver)
Demodulating AM Signals (Using the Analyzer 
As a Fixed Tuned Receiver) 
The zero span mode can be used to recover amplitude modulation on a 
carrier signal. The analyzer operates as a fixed-tuned receiver in zero 
span to provide time domain measurements.
Center frequency in the swept-tuned mode becomes the tuned 
frequency in zero span. The horizontal axis of the screen becomes 
calibrated in time only, rather than both frequency and time. Markers 
display amplitude and time values. 
The following functions establish a clear display of the waveform: 
Trigger stabilizes the waveform trace on the display by triggering on 
the modulation envelope. If the modulation of the signal is stable, 
video trigger synchronizes the sweep with the demodulated 
waveform. 
Linear mode should be used in amplitude modulation (AM) 
measurements to avoid distortion caused by the logarithmic 
amplifier when demodulating signals. 
Sweep time adjusts the full sweep time from 5 ms to 2000 s. (20 µs to 
2000 s if Option AYX is installed). The sweep time readout refers to 
the full 10-division graticule. Divide this value by 10 to determine 
sweep time per division. 
Resolution and video bandwidth are selected according to the signal 
bandwidth. 
Each of the coupled function values remains at its current value when 
zero span is activated. Video bandwidth is coupled to resolution 
bandwidth. Sweep time is not coupled to any other function. 
NOTE
Refer to 
 fo
more information on signal demodulation.
To obtain an AM signal, you can either connect a source to the analyzer 
input and set the source for amplitude modulation, or connect an 
antenna to the analyzer input and tune to a commercial AM broadcast 
station. 
Demodulating an AM Signal Example 1:
View the modulation waveform of an AM signal in the time domain. 
1.
Connect an RF signal source to the analyzer INPUT. For this 
example, an Agilent E4433B Signal Generator was used with the 
following settings: 
a.
RF Frequency 300 MHz