Agilent Technologies E7405A User Manual

Page of 92
Chapter 2
89
Making Complex Measurements
Demodulating and Listening to an AM Signal
7. The signal is demodulated at the marker position only for the 
duration of the demod time. Use the step keys, knob, or numeric 
keypad to change the dwell time. For example, press the step up key 
(
) to increase the dwell time to 2 seconds. 
8. The marker search functions can be used to move the marker to 
other signals of interest. Press 
Peak Search
 to access 
Next Peak
Next Pk Right
, or 
Next Pk Left
Demodulating and Listening to an AM Signal 
Example 2:
1. Connect an antenna to the analyzer input.
2. Perform a factory preset by pressing 
Preset
Factory Preset
 (if 
present). 
3. Select a frequency range on the analyzer, such as the range for AM 
radio broadcasts. For example, the frequency range for AM 
broadcasts in the United States is 550 kHz to 1650 kHz. Press 
FREQUENCY
Start Freq
, 550, 
kHz
Stop Freq
, 1650, 
kHz
.
4. Place a marker on the signal of interest. Press 
Peak Search
 to place a 
marker on the highest amplitude signal, or press 
Marker
Normal
 and 
move the marker to a signal of interest.
5. Set the resolution bandwidth to spectrum analyzer by pressing 
BW/Avg
 and 
Res BW
(SA).
6. Set the frequency of the signal to center frequency by pressing 
FREQUENCY
 then 
Signal Track
 (On) if the signal of interest is the 
highest amplitude on screen signal. If it is not the highest amplitude 
signal on screen, move the signal to center screen by pressing 
Peak Search
, Marker
, and 
Mkr
CF
7. If the signal track function is on, press 
SPAN
, 1, 
MHz
 to reduce the 
span to 1 MHz. If signal track is not used, use the step down key (
to reduce the span and use 
Mkr
CF
 to keep the signal of interest at 
center screen.
8. Set the span to zero by pressing 
SPAN, Zero Span
Zero Span
 turns off 
the signal track function.
9. Change the resolution bandwidth to 100 kHz by pressing 
BW/Avg
Res BW
, then enter 100, 
kHz
.
10.Set the signal in the top two divisions of the screen by changing the 
reference level. Press 
AMPLITUDE
, and then the step down key (
until the signal is in the top two divisions. Set the amplitude scale to 
linear by pressing 
Scale Type (
Lin). Keep the signal center displayed 
by pressing 
AMPLITUDE, Ref Level
 and using the (
) (
) step keys.