Table 2.2-7: Association Reject Reasons
Result
Response
Reason/Diag
The Dicom library refuses the association
SCU Device not known
The Dicom library refuses the association
Maximum number of association 
processing threads exceeded
The Dicom library refuses the association
A799
Dicom Library is not allowed to accept 
C-STORE commands from this device 
and therefore refuses the association.
The Dicom library refuses the association
A0102
SOP Class is not found back in the 
configuration
The Dicom library refuses the association
A702
The Dicom library is unable to create the 
Dicom media file due to the fact that the 
disk is full
The Dicom library refuses the association
A703
The Dicom library is unable to create the 
Dicom media file due to resource 
problems other than ‘disk is full’.
The Dicom library refuses the association
A703
The Dicom library is unable to acquire 
the complete C-STORE request due to 
network problems. 
The Dicom library refuses the association
A701
The Dicom library is unable to acquire 
the complete C-STORE request due to 
network problems (time-out while 
reading data from socket) 
2.2.1.3.1.2
Proposed Presentation Contexts
The DICOM Store AE is capable of proposing the Presentation Contexts shown in the following table:
 
 