AGFA HE/001132 User Manual

Page of 93
HE/001132
Page 36 of 92
Document No. 001132, Revision 2.0
Agfa HealthCare
NodeID Livelink : 14140190
14 June, 2007
2.2.2.3.1.2
Proposed Presentation Contexts
The DICOM Print AE is capable of proposing the Presentation Contexts shown in the following table:
Table 2.2-23: Presentation Contexts Proposed by DICOM Print AE
SOP Class
Transfer Syntax
Extended 
Negotiation
SOP Class
SOP Class UID
Name
UID
Role
Verification 
1.2.840.10008.1.1
DICOM Implicit 
VR Little Endian
1.2.840.10008.1.2
SCU
None
Basic Grayscale Print Management 
Meta SOP Class
1.2.840.10008.5.1
.1.9
DICOM Implicit 
VR Little Endian
1.2.840.10008.1.2
SCU
None
Print Job SOP Class
1.2.840.10008.5.1
.1.14
DICOM Implicit 
VR Little Endian
1.2.840.10008.1.2
SCU
None
Presentation LUT SOP Class
1.2.840.10008.5.1
.1.23
DICOM Implicit 
VR Little Endian
1.2.840.10008.1.2
SCU
None
Note:
The Presentation Context shall use Abstract Syntax IDs which correspond to the SOP Classes UID of 
the Meta SOP Class specified in the first column of the Transfer Syntax Table or included SOP 
Classes. None of the included SOP Classes supports extended negotiation. 
2.2.2.3.1.3
SOP Specific Conformance - Basic Grayscale Print Management
Meta SOP Class
"Workstation" provides Standard conformance to the DICOM Basic Grayscale Print Management Class 
(1.2.840.10008.1.1.9) as SCU.
Support for Basic Grayscale Print Management as SCU also implies support for the following SOP Classes 
as SCU:
1. Basic Film Session SOP Class
2. Basic Film Box SOP Class
3. Basic Grayscale Image Box SOP Class
4. Printer SOP Class
2.2.2.3.1.3.1
Basic Film Session SOP Class (1.2.840.10008.5.1.1.1)
The Basic Film Session IOD describes the presentation parameters which are common for all the films of a 
film session (e.g. number of films, film destination, …)
The Basic Film Session SOP Instance refers to one or more Basic Film Box SOP Instance (§
).
"Workstation"(SCU) can send the following DIMSE services: