AGFA HE/001132 User Manual

Page of 93
HE/001132
Page 39 of 92
Document No.001132, Revision 2.0
NodeID Livelink : 14140190
Agfa HealthCare
14 June, 2007
Attribute Name
Tag
Value
Configuration Information
(2010,0150)
Configurable
Illumination
(2010,015E)
Configurable
Reflective Ambient Light
(2010,0160)
Configurable
Referenced Film Session 
Sequence
(2010,0500)
A sequence which provides references to a Film Session SOP 
Class/Instance pairs. Only a single Item is permitted in this 
Sequence.
>Referenced SOP Class UID
(0008,1150)
Uniquely identifies the referenced SOP Class
>Referenced SOP Instance UID
(0008,1155)
Uniquely identifies the referenced SOP Instance.
Referenced Presentation LUT 
Sequence 
(2050, 0500)
A sequence which provides references to a Presentation LUT 
related SOP Class/Instance pairs. Only a single Item shall be 
included in this sequence.
>Referenced SOP Class UID
(0008, 1150)
Uniquely identifies the referenced SOP Class
>Referenced SOP Instance UID
(0008, 1155)
Uniquely identifies the referenced SOP Instance.
"Workstation" (SCU) will process the N-CREATE confirmation and response Status codes. The status codes 
below are recognized:
Table 2.2-28: N-CREATE Status Codes
Code
Status
Meaning
0000
Success
Normally returned by the SCP AE. Film Box successfully created.
B605
Warning
Requested Min Density or Max Density outside of printer’s operating range. The printer will use its 
respective minimum or maximum density value instead.
*
Other 
Status 
codes
Other Dicom error codes result in the failure of the job. Other warnings are not communicated to 
the user.
2.2.2.3.1.3.3
Basic Grayscale Image Box SOP Class (1.2.840.10008.5.1.1.4)
The Basic Image Box IOD is an abstraction of the presentation of an image and image related data in the 
image area of a film. The Basic Image Box IOD describes the presentation parameters and image pixel data 
which apply to a single image of a sheet of film. 
The Basic Grayscale Image Box SOP Instance is created by the SCP at the time as the Basic Film Box SOP 
Instance (§ 
) is created, based on the value of the Basic Film Box Attribute Image Display 
Format (2010, 0010).
The Basic Grayscale Image Box SOP Instance refers to zero or one Image Overlay Box SOP Instance and 
zero or one Presentation LUT SOP Instance.
"Workstation" (SCU) can send the following DIMSE services:
·
N-SET
N-SET is issued by "Workstation" (SCU) to update an instance of the Grayscale Image Box SOP Class.
When all needed Basic Grayscale Image Boxes have been set, "Workstation" (SCU) issues a print 
command. There can be empty image positions. By using N-SET, "Workstation" (SCU) can instruct the SCP 
to erase the image in the image position by setting a zero length and no value in the attribute Basic 
Grayscale Image Sequence.  The N-Set attributes for the Basic Grayscale Image Box are listed below: