2.4.1.2.2.1.2
Protocol Codes
Note on Japanese Protocol Codes
In Japan, the use of protocol codes is slightly different. See the guidelines of the JIRA and JAHIS [JAPAN].
Each Scheduled Protocol Code Sequence consists of the following information:
Table 2.4-3: Information in a Japanese Scheduled Protocol Code Sequence.
Item
Coding Scheme 
Designator
No. of items
Code Value/Code Meaning
Procedure content
JJ1017T
1
Specified in Section 5.3 
Target region
JJ1017P
1
Specified in Section 5.4 
13
Imaging direction
JJ1017D
0-N
Specified in Section 5.5 
13
(Table copied from [JAPAN], p.9)
Each item mentioned in the table is wrapped in a scheduled protocol code sequence item. As shown in the 
table, for each item a different coding scheme designator is used.
In practice, a triplet (with the imaging direction possibly omitted) refers to a single Exposure Type in the 
"Workstation" exposure tree.
Example of a Japanese SPS
Example of a Scheduled Protocol Code Sequence in a Japanese SPS:
Coding 
Scheme 
Designator
Coding 
Scheme 
Version
Code 
Value
Code Meaning
JJ1017T
1.0
GX.01.00
Radiography – General Radiography – NOS
JJ1017P
1.0
25.6.201
Chest – Respiratory System – Lung
JJ1017D
1.0
G-5200
Anterior to posterior
JJ1017D
1.0
G-A101
Left lateral
This SPS schedules 2 exposures: Chest AP and Chest LL
"Workstation" can be configured for the use of the Japanese system of Protocol Codes (JJ1017) by means of 