Emerson Process Management CFA-3000 User Manual

Page of 74
39
Figure 7­3. Analog Board DIP Switches
(with  #3 set in OFF position for multi­stream analyzer) Colorimeter Assembly
CFA3000 MANUAL
SECTION 7
INSTRUMENT DESCRIPTION
7.3.6.
Analog Board DIP Switches
The following list describes the conditions for each switch to be in the ON or OFF position:
Table 7­9.   Analog Board DIP Switch Positions
Switch No.
Position
Condition
Remarks
1
ON
When using Analog Board, Part 
# 180­B026­02
Analog Board has no adjustment  
ports for individual output device 
and/or 0­5 Vdc, 4­20 mA outputs.
1
OFF
When using Analog Board, Part 
# 180­B026­01
Analog Board has individual ports 
for adjusting each output device 
(up to 6) and/or 0­5 Vdc,
4­20 mA outputs.
2
ON
Quick calibration ­­ on
2
OFF
Quick calibration ­­ off
3
ON
4
ON
Non­functioning switch
5
ON
Non­functioning switch
6
ON
Non­functioning switch
ON
For stream ID and “brackets” before
and after calibration
Allows markers to be recorded on  
strip charts  and other recording  
devices.
7
OFF
No markers will be recorded on  
strip charts and other recording 
devices.
8
ON
For 80 column printers
Formats printer output
8
OFF
For 40 column printers
Formats printer output