Emerson Process Management CFA-3000 User Manual
39
Figure 73. Analog Board DIP Switches
(with #3 set in OFF position for multistream analyzer) Colorimeter Assembly
CFA3000 MANUAL
SECTION 7
INSTRUMENT DESCRIPTION
7.3.6.
Analog Board DIP Switches
The following list describes the conditions for each switch to be in the ON or OFF position:
Table 79. Analog Board DIP Switch Positions
Switch No.
Position
Condition
Remarks
1
ON
When using Analog Board, Part
# 180B02602
# 180B02602
Analog Board has no adjustment
ports for individual output device
and/or 05 Vdc, 420 mA outputs.
and/or 05 Vdc, 420 mA outputs.
1
OFF
When using Analog Board, Part
# 180B02601
# 180B02601
Analog Board has individual ports
for adjusting each output device
(up to 6) and/or 05 Vdc,
420 mA outputs.
(up to 6) and/or 05 Vdc,
420 mA outputs.
2
ON
Quick calibration on
2
OFF
Quick calibration off
3
ON
4
ON
Nonfunctioning switch
5
ON
Nonfunctioning switch
6
ON
Nonfunctioning switch
7
ON
For stream ID and “brackets” before
and after calibration
and after calibration
Allows markers to be recorded on
strip charts and other recording
devices.
devices.
7
OFF
No markers will be recorded on
strip charts and other recording
devices.
strip charts and other recording
devices.
8
ON
For 80 column printers
Formats printer output
8
OFF
For 40 column printers
Formats printer output