Emerson Process Management CFA-3000 User Manual

Page of 74
43
CFA3000 MANUAL
SECTION 7
INSTRUMENT DESCRIPTION
Part
Function
Intake/Regulating Valve
Regulates flow of 
sample into overflow  
sampler assembly
Overflow Tube
Assures adequate supply  
of sample and 
reproducible results by  
eliminating positive
pressure from sample  
stream.
Sample Intake Tubing
Tubing which carries   
sample from overflow  
tube to analyzer (one to  
six stream configurations)
NOTE: Sample intake  
tube should be no more 
than 1 to 1­1/2" into the 
sample.
Air Vent
Allows air to escape  
overflow tube as it fills   
with sample.
Part
Function
Overflow Waste Vent
Allows excess sample
to escape.
Waste Pipe
Carries waste to drain.
Overflow Tubing
All overflow tubes feed  
into common waste pipe.
Filters 
(180­1341­08)
Replaceable filters are  
mounted down stream  
from each intake
regulating valve.
Sample Inlet Tubing
1/8" x 1/4"  tubing 
(provided by customer). 
Common Panel
Mounts multi­stream  
Overflow Sampler  
Assembly to left side of  
analyzer (facing front  
panel).
Panel with dimensions 
22­1/4” x 9" will support  
up to six streams.
7.9:  Overflow Sampling 
Assemblies 
Table 7­15: Overflow Sampling Assembly Parts 
Figure 7­8. Single­Stream Overflow Sampling Assembly