Bacharach 24-9351 Manual De Usuario

Descargar
Página de 100
Instruction 24-9351
PCA
5.3   Calibrate Menu Screen
«TA-ZERO    CALIBRATE
 TA-SPAN     NX
 TL-ZERO     CO
 TL-SPAN     DRAFT
The Calibrate Menu Screen is displayed after entering the correct pass-
word in the Maintenance Password Screen (Section 4.19) and selecting
CALIBRATION from the Maintenance Screen (Section 4.20). Use this
screen to select the sensor to be calibrated.
Press the 
st
 keys until the cursor (
z
) is in front of the desired function,
and then press the 
 key.
NOTE: TA is the stack temperature sensor, while TL is the
optional Room Air/Primary Air temperature sensor.
NOTE:  If a sensor is not installed (i.e., the instrument does
not have a Nitric Oxide sensor or thermocouple simulator
installed), the corresponding calibration screen will not be
displayed.
Front Panel Key Functions:
– Move cursor (
z
) up
– Move cursor (
z
) down
– Select sensor to be calibrated
– Go to Combustion Test Screen
– Return to Maintenance Screen
5-3