ADM800 - Specifications 
Product Description:  The ADM800 Ammonia Display Module provides remote programming, interrogation 
and display functionality to support the AGM300 Ammonia Gas Monitor. The system 
design supports compliance to the refrigerant monitoring requirements of ANS/BSR 
ASHRE 15-1994. 
Inputs: 
The ADM800 accepts inputs from up to four AGM300 monitors. It offers a wide variety 
of displays and can fully program any associated monitor. 
Display: 
Back lit LCD 
Power Indication: 
Green LED glows when AC power is applied 
Alarm Indication:  
Red LED glows and alarm condition is displayed 
Fault Indication: 
Yellow LED glows and fault condition is displayed 
Password Protection:  The ADM800 can be password protected to prevent the unauthorized editing of setup 
parameters without disturbing the ability to monitor system outputs 
Power Safety Mode:  Fully automatic system reset; all parameters maintained 
Monitoring Distance:  Up to 4500 feet from furthest AGM300 
Communications: 
Full two-way communication with AGM300 monitors via RS-485 interface. A second 
RS-485 interface provided for connection to a Building Management System. 
Relays: 
Two SDDT programmable alarm contacts are provided (Rated 5 amp 120 VAC) 
Operating Temp: 
32 to 113 °F (0 to 45 °C) 
Ambient Humidity: 
5 to 90% (non-condensing) 
Size/Weight: 
11"H x 10"W x 3"D - 5 lbs. 
AC Power: 
120 to 230 VAC, 50/60 Hz, 21 Watts 
Mounting: Wall 
mount 
Certification: 
UL 61010-1 and CE Mark 
Warranty: 
Two (2) Years from date of shipment 
Altitude Limit: 2,000 
Meters