Instruction 3015-4275 
53 
Working with the Trend Screen 
Overview - Log Interval 
The AGM300 retains a data log of 100 measurements for each zone. The log interval is the number of minutes 
from 1 to 1440 between each log point. This parameter can be changed from the Zone Setup #1 Screen 
(Page 41). 
The default setting for this parameter is 1440 minutes (24 hours). If the log interval time is set to 0, then a 
measurement is recorded in the trend log after every measurement cycle.  Therefore, after the trend log is filled it 
will contain the last 100 measurement points for a zone.  If you want the data logged less frequently you should 
increase this value. It is important to remember that cycle time is dependent on many factors, including the 
number of zones being monitored, input line length, and the run zeroing mode selected. Before changing this 
value it may be useful to first review the log data using the Trend Screen to determine the nominal cycle time. 
Navigating To the Trend Screen 
Access the Trend Screen from either the System Screen or Zone Screen as follows: 
• From the System Screen (Page 31) press the key adjacent to the AGM300 unit associated with the zone you 
wish to work with to display that monitor's AGM Setup Screen #1. Next, press the ZONES key to go to Zone 
Setup Screen #1
, and then press the TREND key at the bottom of the display to go to the Trend Screen
•  From the top level Zone Screen (Page 44) press the key adjacent to the zone you wish to work with to first 
display its Zone Setup Screen #1. Then press the TREND key at the bottom of the display to go to the Trend 
Screen
Trend Screen 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
The trend graph opens with the cursor located over the most recent data point. Use the LEFT/RIGHT cursor 
keys to move the cursor to different data points. Holding a key down will cause cursor to move rapidly across 
the screen. As you move the cursor position, the date and time of that reading, along with the PPM value, are 
displayed at the top of the screen above the graph. The trend graph is automatically scaled to accommodate 
the largest PPM value displayed. The ZOOM OUT and ZOOM IN keys allow you to adjust the vertical scale of 
the graph. 
The trend data is stored on a first-in, first-out basis.  After 100 trend values have been stored the 101
st
 value will 
replace the first value stored.  Therefore, in normal operation, when entering the trend mode the cursor will be 
located at the most recent data point.  The data points to the left of the initial cursor location will be the next 
most recent.  The data point to the right will be the oldest data point in the buffer and will be over written by the 
next data point.