Working with the Calibration Screen 
Overview 
The Calibration Screen is used to adjust the IR sensor’s calibration factor. 
IMPORTANT:  Changing information on CAL FACTORS will VOID the factory calibration.  Typically, the unit will 
remain within the factory-calibrated accuracy indefinitely and no calibration is required.  Complex software 
algorithms adjust for temperature drift, IR source aging, and pressure changes in order to keep the unit within 
factory accuracy specifications.   
Navigating to the Calibration Screen  
First, display the System Screen (Page 31). Next, press the key adjacent to the target AGM300 unit to go to 
that monitor’s AGM Setup Screen #1 (Page 37). Finally, press the CAL key to go to the selected monitor’s 
Calibration Screen.
 
Calibration Screen 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Adjusting Calibration Factor 
The factory default cal factor for standard units is 1.000.   
Proceed as follows to adjust the current calibration factor: 
1.  Press the CAL FACTOR key to edit the value. 
2.  Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to 
modify the individual numbers. 
3.  Press ENTER to accept the new entry or QUIT to revert to the previous setting.