Bacharach AGM300 Manual De Usuario

Descargar
Página de 94
Instruction 3015-4275 
65 
Optional Accessories 
Intake Manifolds 
Up to three additional intake manifolds (P/N 3015-3419 & P/N 3015-3411) can be installed in the AGM300 to 
increase the total capacity to 16 zones. Each manifold requires a sample line-end filter for each of the 4 zones. 
Please follow the instructions provided to install the manifolds. Then follow the instructions earlier in this manual 
to install the air lines and complete the zone programming steps. The AGM300 setup must also be updated to 
recognize the new zones. 
4–20 mA Interface Board 
The optional 4–20 mA Interface Board (P/N 3015-3133) (Page 7) allows the operator to take advantage of the 
current loop interfaces provided for connection of external devices to the AGM300. The interface board plugs 
into the socket provided on the main circuit board, and the external devices are then cabled to the AGM300 as 
described on Page 16.