Instruction 3015-4275 
79 
Service Mode Register  
Register 0x001Bh 
10 bytes 
Variable Type 
Length 
Description 
Rel_Svc_Mode *  * 
See description of STATUS REGISTER 
 
 
Release Service Mode Register 
Register 0x001Ch 
10 bytes 
Variable Type 
Length 
Description 
Ent Svc_Mode 
See description of STATUS REGISTER 
 
 
 
 
Notes Concerning the AGM300 Service Mode 
The AGM300 can be placed into service mode if necessary.  During service mode the unit will take no 
measurements, any and all alarms are silenced, and all relays are opened.   The unit automatically comes out 
of service mode after a preset interval defined by the service_mode_TO parameter in the System Data 
Register.  In order to place the AGM300 into service mode use the following sequence: 
•  Read the AGM300 Status Register (0x0011h) 
•  Modify the content of the status register structure to change the MODE parameter to service mode. 
•  Send this updated status register structure back to the AGM300 using PRESET MULTIPLE REGISTER 
COMMAND to the SERVICE MODE REGISTER (0x001Bh). 
 
To Release the Unit from Service Mode Use the  
Following Sequence of Steps 
•  Read the AGM300 Status Register (0x0011h) 
•  Modify the content of the status register to change the MODE parameter to normal mode. 
•  Send this updated status register structure back to the AGM300 using PRESET MULTIPLE REGISTER 
COMMAND to the RELEASE SERVICE MODE REGISTER (0x001Ch).