90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 33
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Linearization
↓↓
In the menu "Linearization" and its submenus you can set up and start the calculation to
linearize one channel of an MLT analyzer or analyzer module. With the F3 key you can
change to the other channels. With the F5 key (Info...) you can change to the submenu
"Linearization Info", where you will find the results and some parameters of linearization.
Background:
The optical absorption of a gas as a function of its concentration is not linear in the whole
range. Therefore the result of the analyzer has to be linearized. You can linearize manual
or semi-automatically. In this chapter the manual linearization is described. If you want to
linearize automatically, contact your customer service, please.
Measure
Back...
Info...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Linearization --
Linearization Procedure...
Linearization setpoint values...
Set all raw/setpoint values to 0.0 !
Calculate linearization curve coefficients...
Linearization raw values...
Linearization is:
Disabled
Channel
Measure
Channel
Back...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Linearization Info --
Linearization is:
Disabled
Method in use:
Splines
Status for calculation of 4-order polyn. method:
Ready
Status for calculation of spline method:
Ready
Lower linearization limit is:
-200000.00  ppm
Upper linearization limit is:
1200000.0  ppm
Linearization is:
In Range
5.1 Analyzer Module Setup
5.1.5 Linearization
In these lines
you can control
if the calculated
linearization
was succesful !