90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 73
Configuration of Relay Outputs:
Press the 
88 
 -key or the 
 -key in the line "Configuration of Relay Outputs..." of the menu
"Local SIO Configuration Parameters" to change to the corresponding menu:
In the menu "Relay Outputs" you can attach signals to the relay outputs of the local SIO
board, e.g the relation between the zero valve and a certain channel. You have to put a
number code (= variable of a signal) in the corresponding line for the relay:
There are three relays on the local SIO board. With a jumper you can determine the relay
contact status: NO or NC. Informations about the whole specification of the SIO board you
can find in another manual.
Note:
In this menu you only can determine the relay output configuration for channels of the
analyzer module, not for channels of auxiliary modules.
Set up parameters:
♦ 
Select any line of variables with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable with the 
88 
 -key or the 
 -key.
♦ 
Select the code number for the signal with the 
↑↑
 -key or the 
↓↓
 -key.
+
+
♦ 
Confirm the new value with the 
88 
 -key or
cancel and go back to the last value with the F2 key.
Lines of variables "Relay 1, 2, 3 - actual state":
These lines display the current gear status of each relay. The lines are not available.
Measure
Back...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Relay Outputs --
Relay 1 - output signal:
4
Relay 3 - output signal:
6
Relay 2 - actual state:
Off
Relay 3 - actual state:
Off
Relay 2 - output signal:
5
Relay 1 - actual state:
On
5.1 Analyzer Module Setup
5.1.18 Inputs and Outputs – Local SIO