Siemens SITRANS FUS1010 NEMA-4X/ IP 65 Manual De Usuario

Descargar
Página de 232
      3-2
1010NFM-3J
Section 3
z
Avoid mounting the transducers on a section of pipe with any external scale. Remove all scale,
rust, loose paint, etc., from the location.
z
Do not mount the transducers on a surface aberration (pipe seam, etc.)
z
Do not mount transducers from different ultrasonic flowmeters on the same pipe. Also, do not run
the transducer cables in common bundles with cables from communication equipment, other Con-
trolotron systems, or any type of ultrasonic equipment. You can run these cables through a com-
mon conduit ONLY if they originate at the same flowmeter.
z
Never mount transducers under water, unless you order submersible units and you install them in
accordance with factory instructions.
z
Never mount transducers on the top or bottom of a horizontal pipe. The best placement on a
horizontal pipe is either the nine o’clock or three o’clock position for Reflect Mode, or one trans-
ducer at nine o’clock and one transducer at three o’clock for Direct Mode. Mounting on a vertical
pipe is recommended only if flow is in the upward direction.
When mounting on a vertical pipe flowing in a downward direction make sure there is sufficient back
pressure in the system to maintain a full pipe.
3.1.3 CLAMP-ON TRANSDUCER MOUNTING MODES
Siemens clamp-on transducers support Direct or Reflect mounting modes. The flow computer recom-
mends a mounting mode after analyzing your pipe and liquid data entries. However, you can install
clamp-on transducers in the way that best suits your application and the transducer type you pur-
chased.
REFLECT MOUNT
(Pipe shown in Top View for ease in visualizing sonic path)
Upstream
Transducer
Sonic Path
Pipe
Downstream
Transducer
END VIEW
(Horizontal Plane)
Transducer locate d
at 9  o'clock
Transducer locate d at
3 o'clock