ACCESSORIES
pH CALIBRATION SOLUTIONS
HI 70004P pH 4.01 Buffer Sachets, 20mL, 25 pcs
HI 70007P pH 7.01 Buffer Sachets, 20mL, 25 pcs
HI 70010P pH 10.01 Buffer Sachets, 20mL, 25 pcs
HI 7004L
pH 4.01 Buffer Solution, 460 mL
HI 7006L
pH 6.86 Buffer Solution, 460 mL
HI 7007L
pH 7.01 Buffer Solution, 460 mL
HI 7009L
pH 9.18 Buffer Solution, 460 mL
HI 7010L
pH 10.01 Buffer Sol., 460 mL
ELECTRODE STORAGE SOLUTION
HI 70300L Storage Solution, 460 mL
ELECTRODE CLEANING SOLUTIONS
HI 70000P Electrode Cleaning Sachets, 20 mL, 25 pcs
HI 7061L
General Cleaning Sol., 460 mL
HI 7073L
Protein Cleaning Sol., 460mL
HI 7074L
Inorganic Cleaning Sol., 460mL
HI 7077L
Oil & Fat Cleaning Sol.,460 mL
REFILL ELECTROLYTE SOLUTIONS
HI 7071
3.5M KCl + AgCl Electrolyte, 4x50 mL
HI 7072
1M KNO
3
 Electrolyte, 4x50 mL
HI 7082
3.5M KCl Electrolyte, 4x50 mL, for double junction electrodes
electrodes only): refill with fresh HI7071  for single junction or
HI7082 for double junction electrodes.
• Dry Membrane/Junction: Soak in Storage Solution HI70300
for at least 1 hour.
• Drifting: Soak the electrode tip in warm Hanna Solution
HI7082 for one hour then flush tip with distilled water. Refill with
fresh HI7071 for single junction electrodes and HI7082 for double
junction electrodes.
• Low Slope: Refer to the cleaning procedure above.
• No Slope: Check the electrode for cracks in glass stem or bulb
and replace the electrode.
• Slow Response/Excessive Drift: Soak the tip in Hanna Solu-
tion  HI7061  for 30 minutes, rinse thoroughly in distilled water
and then follow the Cleaning Procedure above.