Instruction Manual 
 
Chloride Electrode 
 
14
The time needed to reach 99% of the stable electrode potential reading, the electrode response time, 
varies from several seconds in highly concentrated solutions to several minutes near the detection 
limit.  (Refer to Figure 2.) 
 
A drifting potential reading or a decrease in electrode slope may mean that the electrode membrane 
needs polishing. 
 
To polish the membrane: 
 
1. 
If using polishing paper, cut off a 1-2" piece and place it face up on the lab bench. 
 
2. 
Put a few drops of distilled or deionized water in the center of the paper. 
 
3. 
Holding the paper (cotton) steady with one hand, bring the membrane of the electrode 
down perpendicular to the paper and, with a slight swirling motion, gently polish the tip of 
the electrode against the surface of the polishing paper (cotton) for a few seconds. 
 
 
4. 
Rinse the electrode surface with distilled or deionized water and soak the electrode tip in 
standard solution for about five minutes before use. 
 
5. 
If using jeweller's rouge, place a cotton ball on the table top and flatten it using the bottom 
of a beaker. 
 
6. 
Put 1-2 drops of distilled or deionized water in the center of the cotton pad. Add a small 
amount of jeweller's rouge to the damp cotton. 
 
7. 
Continue with Steps 3 and 4 above.