 PK-7000-MPC1-D Interconnections Kit
 
 
 
PG7000, RPM, PPC and MPC are trademarks, registered and otherwise, of DH Instruments, Inc.
 
Document 560039 
010509 
DH Instruments, Inc. 
4765 East Beautiful Lane 
Phoenix  AZ  85044-5318 
USA 
 
 
Tel 602.431.9100
Fax 602.431.9559
dhi@dhinstruments.com
www.dhinstruments.com 
 
Screw 1/8 in. tube x 1/8 in. NPT male adaptor into MPC1 TEST port using Teflon tape as a seal. 
ra   To use quick connect stand for TEST connection:  Install reducing union into quick connecting head.  
Mount quick connect stand on bench using mounting holes provided. 
rb   To use 1/4 in. NPT female adaptor for TEST connection:  Install 1/8 in. tube x 1/4 in. NPT female 
adaptor onto ends of tube from MPC1-D TEST port (Step 
u below). 
Screw 1/8 in. tube x 1/8 in. NPT male adaptor into MPC1 REF port using Teflon tape as a seal.  
Connect PG7000 or RPM3 to MPC1-D REF port by cutting a piece of 1/8 in. SS tubing to desired 
length and swage together. 
Connect quick connect stand or 1/8 in. tube x 1/4 in. NPT female adaptor to MPC1-D TEST port by 
cutting a piece of 1/8 in. SS tubing to desired length and swage together. 
 
• 
1/8 in. SS tubing can be bent by hand to desired shape but use caution not to kink the tube.