Instructions for the Physician
Implanted Stimulation Devices.
If other implanted devices are
indicated for the patient, careful screening is required to determine if
safe results can be achieved before permanently implementing con-
current electrical therapies.
safe results can be achieved before permanently implementing con-
current electrical therapies.
Postural Changes.
Depending on the activity level of the patient,
postural changes may affect stimulation intensity. Instruct patients to
keep the Remote Control at hand at all times, and ensure that they
understand how to adjust stimulation levels.
keep the Remote Control at hand at all times, and ensure that they
understand how to adjust stimulation levels.
Medical Devices/Therapies.
If the patient is required to undergo
lithotripsy, electrocautery, external defibrillation, radiation therapy,
ultrasonic scanning, or high-output ultrasound:
ultrasonic scanning, or high-output ultrasound:
•
Turn off stimulation at least five minutes before the pro-
cedure or application.
cedure or application.
•
All equipment, including ground plates and paddles,
must be used as far away from the IPG as possible.
must be used as far away from the IPG as possible.
•
Bipolar electrocautery is recommended: Do not use
monopolar electrocautery.
monopolar electrocautery.
•
Every effort should be taken to keep fields, including
current, radiation, or high-output ultrasonic beams,
away from the IPG.
current, radiation, or high-output ultrasonic beams,
away from the IPG.
•
Equipment should be set to the lowest energy setting
clinically indicated.
clinically indicated.
•
Instruct patients to confirm IPG functionality following
treatment by turning on the IPG and gradually increas-
ing stimulation to the desired level.
treatment by turning on the IPG and gradually increas-
ing stimulation to the desired level.