Physician Lead Manual
Instructions for the Physician
Implanted Stimulation Devices. 
If other implanted devices are 
indicated for the patient, careful screening is required to determine if 
safe results can be achieved before permanently implementing con-
current electrical therapies.
Postural Changes. 
Depending on the activity level of the patient, 
postural changes may affect stimulation intensity. Instruct patients to 
keep the Remote Control at hand at all times, and ensure that they 
understand how to adjust stimulation levels.
Medical Devices/Therapies. 
If the patient is required to undergo 
lithotripsy, electrocautery, external defibrillation, radiation therapy, 
ultrasonic scanning, or high-output ultrasound:
Turn off stimulation at least five minutes before the pro-
cedure or application.
All equipment, including ground plates and paddles, 
must be used as far away from the IPG as possible.
Bipolar electrocautery is recommended: Do not use 
monopolar electrocautery.
Every effort should be taken to keep fields, including 
current, radiation, or high-output ultrasonic beams, 
away from the IPG.
Equipment should be set to the lowest energy setting 
clinically indicated.
Instruct patients to confirm IPG functionality following 
treatment by turning on the IPG and gradually increas-
ing stimulation to the desired level.