Braemar Manufacturing LLC FUSION Manual De Usuario

Descargar
Página de 16
 
 
 
 
      
    Fusion Wireless Recorder 
 
21 
Electromagnetic Emissions 
Emissions test  Compliance 
Electromagnetic  
environment - guidance 
RF emissions 
CISPR 11 
Class B 
Fusion is suitable for use in all 
establishments, including domestic 
establishments and those directly 
connected to the public low-
voltage power supply network that 
supplies buildings used for 
domestic purposes. 
 
Electromagnetic Immunity 
Immunity test 
IEC 60601 
test level 
Compliance 
level 
Electromagnetic 
environment - guidance 
Electrostatic 
discharge (ESD)  
IEC 61000-4-2 
±6 kV contact 
±8 kV air 
±6 kV contact 
±8 kV air 
Floors should be wood, 
concrete or ceramic tile. If 
floors are covered with 
synthetic material, the relative 
humidity should be at least 
30% without condensation. 
Power frequency 
(50/60 Hz) 
magnetic field 
IEC 61000-4-8 
3 A/m 
3 A/m 
Power frequency magnetic 
fields should be at levels 
characteristic of a typical 
location in a typical 
commercial or hospital 
environment. 
 
 
 
 
 
 
      
    Fusion Wireless Recorder 
 
22