Braemar Manufacturing LLC FUSION Manual De Usuario
Fusion Wireless Recorder
9
Recorder Preparation
1. General setup: Remove the Patient Cable if it is connected to
the Recorder. Open the battery compartment by sliding battery
door upward. Install the AA Lithium battery pack. Observe proper
battery polarity. The Recorder will sound rising tones after
completion of power up. After a few seconds, the splash screen will
appear. If a patient cable is not connected, a message indicating that
a cable is not connected will be displayed. Please insert a patient
cable to exit the splash screen and enter the Setup (New Recording)
Menu with the battery level in the upper right corner. The battery
level should be near 100%. All button presses should beep to
provide feedback for the user. The backlight for the display is on
while accessing the Menus, after recording a manual event, or after
receiving a text message. The backlight will remain off during
normal recording.
door upward. Install the AA Lithium battery pack. Observe proper
battery polarity. The Recorder will sound rising tones after
completion of power up. After a few seconds, the splash screen will
appear. If a patient cable is not connected, a message indicating that
a cable is not connected will be displayed. Please insert a patient
cable to exit the splash screen and enter the Setup (New Recording)
Menu with the battery level in the upper right corner. The battery
level should be near 100%. All button presses should beep to
provide feedback for the user. The backlight for the display is on
while accessing the Menus, after recording a manual event, or after
receiving a text message. The backlight will remain off during
normal recording.
2. Erase Data:
If the Recorder still has data and the Recording is Complete, you
must erase all the data before setting up the Recorder for the next
patient.
must erase all the data before setting up the Recorder for the next
patient.
A. Remove the Patient Cable.
B. Press the Left and Right buttons together and it will prompt you
to Erase Data.
C. Press the Enter button to erase the data. The display will have
the message “Recorder Empty”.
D. Insert the patient cable, press the Left and Right buttons
together, and the Setup (New Recording) Menu will appear with
the battery level in the upper right corner. The battery level
should be near 100%.
the battery level in the upper right corner. The battery level
should be near 100%.
Fusion Wireless Recorder
10
If the Recorder is still recording data, you must stop the recording
and erase all the data before setting up the Recorder for the next
patient.
and erase all the data before setting up the Recorder for the next
patient.
A. Remove the Patient Cable.
B. Press the Left and Right buttons together and it will prompt you
if you want to Stop Recording.
C. Press the Enter button to stop recording the data and the
Recording Complete screen will appear.
D. Continue to erase the data by following the steps above.
3. Preliminary Setup:
At this time, the patient leads should be connected to the electrodes
and the electrodes should be connected to the patient.
and the electrodes should be connected to the patient.
Using the left and right arrow buttons, select the “
” to check the
lead status then press the “
” button, also called Enter, to select the
menu option. Any leads are disconnected will have a caution
symbol (“
symbol (“
”) next to it and a “Lead Error” will be displayed.
Also, the “
” changes to “
” when there is lead loss in the
Current Menu icon and in the Setup (New Recording) Menu. At
least two leads must be connected to see the Lead Status. If there
are less than two leads present, a “No Cable” message is displayed.
Press the Enter button to return to the Setup (New Recording) Menu.
least two leads must be connected to see the Lead Status. If there
are less than two leads present, a “No Cable” message is displayed.
Press the Enter button to return to the Setup (New Recording) Menu.
Lastly, use the left and right buttons to select the View ECG option,
“
“
”, and press Enter to check the ECG measurements. You can