Boston Scientific Corporation CRM314013 Manual De Usuario
Proximity to other equipment
Do not use the ZWT adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, check the ZWT for normal operation in that
confi guration.
Do not use the ZWT adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, check the ZWT for normal operation in that
confi guration.
ZWT and PRM must remain outside sterile fi eld
The ZWT and the PRM are non-sterile and cannot be sterilized. They must both remain outside the sterile fi eld. Maintain a minimum operating
distance of 7.6 cm (3 in) between the ZWT and the PRM.
The ZWT and the PRM are non-sterile and cannot be sterilized. They must both remain outside the sterile fi eld. Maintain a minimum operating
distance of 7.6 cm (3 in) between the ZWT and the PRM.
ZWT is MR Unsafe
The ZWT is MR Unsafe and must be kept outside MRI sites classifi ed Zone III and above, as defi ned by the American College of Radiology Guidance
Document for Safe MR Practices . Under no circumstances should the ZWT be brought into the MRI scanner room, the control room, or MRI site Zone III
or IV areas.
The ZWT is MR Unsafe and must be kept outside MRI sites classifi ed Zone III and above, as defi ned by the American College of Radiology Guidance
Document for Safe MR Practices . Under no circumstances should the ZWT be brought into the MRI scanner room, the control room, or MRI site Zone III
or IV areas.
Modifi cations
No modifi cation of this equipment is allowed unless approved by Boston Scientifi c.
No modifi cation of this equipment is allowed unless approved by Boston Scientifi c.
PRECAUTIONS
Connect only to Boston Scientifi c Model 3120 PRM
Connect the ZWT only to a Model 3120 Boston Scientifi c PRM equipped with current level software.
Connect the ZWT only to a Model 3120 Boston Scientifi c PRM equipped with current level software.
Electrical and magnetic interference
Avoid establishing telemetry communication between the ZWT and the pulse generator when the ZWT is in close proximity to monitors,
high-frequency electrosurgical equipment, or strong magnetic fi elds. The telemetry link may be impaired.
Avoid establishing telemetry communication between the ZWT and the pulse generator when the ZWT is in close proximity to monitors,
high-frequency electrosurgical equipment, or strong magnetic fi elds. The telemetry link may be impaired.
2