GDS-2000A Series User Manual 
108
 
XY Mode 
Cursors can be used with XY mode. 
See the Cursor chapter for details. 
Set the Sampling Mode 
 
Background 
The GDS-2000A has two types of sampling modes: 
ET (Equivalent Time) and Sin(x)/x interpolation. 
Equivalent time sampling is able to achieve a 
sample rate of 100GSa/s when sampling periodic 
waveforms. Sin(x)/x interpolation uses a sinc 
interpolation formula to reconstruct a continuous 
signal between sampled points.  
 
Sin(x)/x 
One sample of data is used to 
reconstruct a single waveform. 
Sin(x)/x sampling should be used 
when the time base is relatively 
slow or if single shot events need to 
be captured. 
 
Equivalent-
time sampling 
Sampled data is accumulated a 
number of times to reconstruct a 
single waveform. This increases the 
sampling rate, but can only be used 
for repetitive signals. This mode is 
usually used when the time base is 
too fast for real-time sampling. 
Panel Operation 
1.  Press the Acquire key. 
Acquire
 
 
2.  Press the ET/sin(x)/x key on the 
bottom menu to toggle between 
equivalent time sampling (ET) and 
sin(x)/x interpolation.