E40.0.01.6C-02 Operating 
Manual 
EASY
LOG
 80 CL 
Page 10 of 12 
10.  Specifications 
 
Measuring range:
 
 
 Temperature 
-25.0 … +60.0 °C 
 Humidity  
0.0 … 100.0 %RH 
(recommended range : 10 … 90 %RH) 
 Air pressure 
300.0 … 1100.0 hPa (mbar) 
Additional selectable 
variables  
 
 Wet-bulb temperature 
-27.0 … +60.0 °C 
  Dew point temperature 
-40.0 … +60.0 °C 
 Enthalpy 
-25.0 … 999.9 kJ/kg 
 
Moisture content of the air 
0.0 … 640.0 g/kg 
Resolution 
0.1 °C  /  0.1 %RH /  0.1 hPa 
Accuracy (± 1 digit): 
 
 Temperature 
±0.3 °C ±0.017 * (T – 25 °C) 
 Humidity 
±2 %RH (at range 10 … 90 %RH) 
 Air pressure 
±1.0 mbar (typ.), ±2.5 mbar (max.) 
Display 
two 4½-digit LCD-displays 
(about 7 and 5.8 mm high) 
Interface 
EASYBus 
 Busload 
equals to 2 EASYBus-elements 
Battery life time 
depending on the chosen measuring cycle 
about 5 years 
(at 15 minutes and nominal temperature). 
Measured value memory  250.000 measuring values of each variable
 
Measuring cycle 
4 seconds to 5 hours (18000 seconds) 
Memory type: 
 
 ›cyclic logger‹ 
•  recording can be activated by start/stop function 
(i.e. key-press, boundaries). 
•  64 independent recording sets are possible. 
 ›ring logger‹ 
The logger records continuously; 
as soon as the whole memory is used the old data 
of the beginning are overwritten. 
 
 
 
DIF
hPa
°C
%
°C
MAX
MIN
SET
1
2
3