Gw Instek GPT-9602 Insulation measuring device, GPT-9602 Manual De Usuario

Los códigos de productos
GPT-9602
Descargar
Página de 75
 
OPERATION 
37 
Setting the Test Time (Timer) 
 
Background 
The TIMER setting is used to set the test time 
for the test. The test time determines how long 
to run the test at the test voltage. This test time 
does not include the Ramp up time (fixed at 
100ms) or initial start time. The test time can be 
set from 001 seconds to 180 seconds for ACW, 
DCW and IR tests, with a resolution of 1 second 
for all modes.  The timer can also be turned off. 
The total discharge time depends on the DUT 
and is not specified. 
Discharge time 
time
Output V
TEST TIME
RAMP 100ms
 
Steps 
1.  When in the EDIT status, press the 
FIELD key to bring the cursor to 
the T setting. 
FIELD
 
 
 
     
 
 
 
H = 2 2 . 0 m A  
 
 
 
 
m A
L
0
=
0 0
.
   
7 4
S  
 
 
 
 
 
=
1
T
       
   
CW
D
A CW
R  
I
D
O E
M
T M E
I
cursor
 
1 00
 
k V
I
D
E
 
T
A
m
 
 
 
2.  Use the scroll wheel to set the T 
value*.  
 
 
 
Range 
001s~180s