GPT-9000 Series User Manual
18
GPT-9000 Rear Panel
SIGNAL I/O USB A port RS232 port
Fan
Fuse selector
Line voltage
Optional GPIB port
GND
TO AVOID ELECTRIC SHOCK THE POWER CORD
PROTECTIVE GROUNDING CONDUCTOR MUST BE
PROTECTIVE GROUNDING CONDUCTOR MUST BE
ONLY WITH SPECIFIED TYPE AND RATED FUSE
.
NO OPERATOR SERVICEABLE COMPONENTS INSIDE
.
DO NOT REMOVE COVERS REFER SERVICING TO
FOR CONTINUED FIRE PROTECTION
.
REPLACE
CONNECTED TO GROUND .
QUALIFIED PERSONNEL .
WARNING
AC
LINE VOLTAGE
100V
230V
220V
120V
SELECTION
207~250V
RANGE
90~110V
198~242V
108~132V
(50/60 Hz)
T 5A
250V
250V
FUSE
POWER MAX.
500 VA
GND
SIGNAL I / O
RS 232
SER. NO .
LB
ENSURE THE POWER IS REMOVED FROM
THE INSTRUMENT BEFORE REPLACING THE FUSE
GPIB
T 2.5A
250V
250V
SIGNAL I/O port
SIGNAL I / O
The SIGNAL I/O port is used to
monitor the tester status (PASS,
FAIL, TEST) and input (START/
STOP signals). It is also used with
the Interlock key.
monitor the tester status (PASS,
FAIL, TEST) and input (START/
STOP signals). It is also used with
the Interlock key.
USB A port
Used for remote control.
RS232 interface
port
port
RS232
Used for remote control and
firmware updates.
firmware updates.
Fan
Exhaust fan.
GND
GND
Connect the GND (ground)
terminal to the earth ground.
terminal to the earth ground.
Line voltage input
220
23
0
100
12
0
Line voltage input:
100/120/220/230VAC ±10%
100/120/220/230VAC ±10%