Temperature Measurement: Temp RTD and Temp TC
Temperature measurement is performed with a Pt100 or Pt1000 
resistance thermometer, or a type K thermocouple (accessory, 
not included), which is connected to the voltage input.
Selecting the Unit of Measure for Temperature
 
1nFo 
Z
 ... 
Z
 
tEMP 
 °C / °F 
UV
 
(°C = default setting)
5.3.1 Measurement with Resistance Thermometers
➭ Set the rotary switch to “Temp
RTD
”.
Press the FUNC | ENTER key in order to change to the other 
measuring function if required.
The sensor type, i.e. Pt100 or Pt1000, is detected automatically 
and displayed.
There are two different ways to compensate for cable resistance:
Automatic Compensation
➭ Acknowledge by pressing the ZERO | ESC key.
“Short leads” appears at the display.
If you prefer to enter cable resistance directly, you can skip the 
following entry prompt.
➭ Short circuit the measuring instrument’s connector cables. 
000.0” appears at the display. After pressing the FUNC | 
ENTER key, automatic compensation of cable resistance is 
activated for all subsequent measurements. The short-circuit 
can now be eliminated, and the device is ready for use.
Enter Cable Resistance
➭ Press the ZERO | ESC key once again in the automatic 
compensation menu.
➭ Enter the known resistance of the connector cables with the 
scroll keys: 
Select the digit to be changed with the 
Y
 
Z
 keys, and change 
the respectively selected digit with the 
VU
 keys. The default 
value is 0.1 
Ω. Values can be selected within a range of 0 to 
50
Ω.
➭ Upon pressing the FUNC | ENTER key, the selected value is 
activated and the display is returned to the measuring function. 
Cable resistance remains in memory even after the instrument 
has been switched off.