Gossen Metrawatt Metrahit Iso Insulation measuring device, CAT III 600 V Metrahit Iso Manual De Usuario

Los códigos de productos
Metrahit Iso
Descargar
Página de 72
GMC-I Gossen-Metrawatt GmbH
43
Measurements: V/Hz, 
Ω,  , Temp (°C/°F), 
, M
Ω
@UISO
, A/Hz
5.8.1 Direct and Pulsating Current Measurement, Direct Connection, 
A DC and A (DC+AC)
➭ First disconnect supply power from the measuring circuit or 
the power consumer (1), and discharge any capacitors.
➭ In accordance with the current to be measured, set the rotary 
switch to A   or A
 
.
➭ Select the type of current appropriate for the measured 
quantity by briefly pressing the FUNC | ENTER multifunction 
key. Each time the key is pressed, the instrument is switched 
back and forth between A DC and A (DC + AC)
TRMS
, which is 
indicated by means of an acoustic signal. The current type is 
indicated at the LCD by means of the DC or the (DC+AC)
TRMS
 
symbol.
➭ Safely connect the measuring instrument (without contact 
resistance) in series to the power consumer (2) as shown.
➭ Switch supply power to the measuring circuit back on (3).
➭ Read the display. Make a note of the measured value if the
instrument is not being operated in the memory mode or the 
transmission mode.
➭ Disconnect supply power from the measuring circuit or the 
power consumer (1) once again, and discharge any 
capacitors.
➭ Remove the test probes from the measuring point and return 
the measuring circuit to its normal condition.
COM mA A
V
Temp
Ω MΩ
ISO
~
R
x
~
R
x
1
3
2
!
A
A
0
03.50
mA
DC
0
03.50
mA
DC
AC
TRMS
A
A
A
: 0.1 
μA . . . 11 A
A
: 10 
μA . . . 11 A
6 ranges: 300 
μA, 3 mA, 30 mA, 300 mA, 3 A and 10 A
8
8.8.8.8
!
I > 11 A
Current
Current
Measuring Ranges:
Current measurement 
may only be performed 
with installed batteries!