Maintenance – Calibration
9
Maintenance and Calibration
Attention!
!
Disconnect the instrument from the measuring circuit before 
opening the battery compartment lid or fuse cover when 
replacing batteries or fuses!
9.1
Displays – Error Messages 
9.2
Batteries
Note

Removing the Batteries During Periods of Non-Use 
The integrated quartz movement draws power from the 
batteries even when the instrument is switched off. It is 
advisable to remove the batteries during long periods of 
non-use for this reason (e.g. vacation). This prevents 
excessive depletion of the battery, which may result in 
damage under unfavorable conditions.
Note

Battery Replacement
Stored measurement data are lost when the batteries are 
replaced. In order to prevent data loss, it is advisable to 
backup your data to a PC with the help of METRAwin 10 soft-
ware before replacing the batteries. 
The selected operating parameters remain in memory, 
although date and time must be reentered.
Battery
The current battery charge level can be queried in the “
1nfo” 
menu: 
 
1nFo 
 
bAtt: 2.75 V.
Make sure that no battery leakage has occurred before initial 
start-up, as well as after long periods of storage. Continue to 
inspect the batteries for leakage at short, regular intervals.
If battery leakage has occurred, carefully and completely clean 
the electrolyte from the instrument with a damp cloth, and replace 
the battery before using the instrument.
If the “
” symbol appears at the display, the batteries should 
be replaced as soon as possible. You can continue working with 
the instrument, but reduced measuring accuracy may result.
The instrument requires two 1.5 V batteries in accordance with 
IEC LR 6, or two equivalent rechargeable NiCd batteries.
Message
Function
Meaning
FUSE
Current measurement
Blown fuse
In all operating modes
Battery voltage has fallen below 1.8 V
0L
Measurement
Indicates overflow
ur
M
Ω
ISO
 measurement
Measured value of less than 10% of the mea-
suring range
error
M
Ω
ISO
 measurement
Interference voltage detected
MEASURE