Maintenance – Calibration
11.1
Housing Maintenance
No special maintenance is required. Keep outside surfaces clean and dry. Use a slightly 
dampened cloth for cleaning. Avoid the use of solvents, cleansers and abrasives.
Note!
If the test case has not been used for a long period of time, the switches may 
demonstrate increased contact resistance depending upon storage conditions. 
If this is the case, actuate the switches several times.
11.2
Calibration
According to DIN VDE 0701-0702, measuring instruments used for periodic testing must be 
tested and calibrated in accordance with the manufacturer’s specifications on a regular 
basis. Depending upon usage, the manufacturer recommends an interval of 1 to 3 years for 
this test instrument.
Note!
These tests can be performed on-site with the SECU-cal 10 calibration adapter 
accessory.
11.3
Periodic Self-Test of the Connector Cable for Protective Conductor Continuity
Connect the probe cable (21) to a grounding contact which has been previously tested for 
absence of voltage (e.g. at an electrical outlet), and which is connected to the protective 
conductor in the connector cable, and set the NETZ-VDE switch (1) to “NETZ”. Then 
measure protective conductor resistance as described in section 5.4. If an excessively high 
protective conductor resistance value is displayed at the LCD panel or if overloading is 
indicated (only “O.L” appears), protective conductor resistance is too high or the protective 
conductor is interrupted. Eliminate the interruption (in the cable or at the NETZ-VDE switch).
11.4
Testing the Integrated RCD (residual current circuit breaker)
Test on a regular basis. The integrated RCD (residual current circuit breaker) can be tested 
by pressing the test key. Breaking current value and time can be measured with test 
instruments for DIN VDE 0413, part 6.
11.5
Fuse Replacement
All fuses can be accessed from outside. Only fuses with the breaking characteristics and 
rated current values specified on the front panel may be used.
11.6
Return and Environmentally Sound Disposal
The METRATESTER 5+
3P
 is a category 9 product (monitoring and control instrument) in 
accordance with ElektroG (German electrical and electronic device law). This device is not 
subject to the RoHS directive.
In accordance with WEEE 2002/96/EG and ElektroG, we identify our electrical 
and electronic devices (as of Aug. 2005) with the symbol in accordance with 
DIN EN 50419 which is shown at the right. Devices identified with this symbol 
may not be disposed of with the trash.
Please contact our service department regarding the return of old devices (see 
address below).