Gw Instek GDS-2074A 4-channel oscilloscope, Digital Storage oscilloscope, GDS-2074A Hoja De Datos

Los códigos de productos
GDS-2074A
Descargar
Página de 261
 
 
GDS-2000A Series User Manual 
230
 
USB/RS-232C Functionality Check 
 
Terminal 
Application 
(USB/RS-232C)  
Invoke a terminal application such as RealTerm.  
For RS-232C and USB, set the COM port, baud 
rate, stop bit, data bit, and parity accordingly. 
To check the COM port number and associated 
port settings, see the Device Manager in the PC. 
For WinXP: 
 Control panel → System → Hardware tab 
Example: Configuring RealTerm for RS232C 
communication. 
 
Functionality 
Check 
Key in this query command via the terminal 
application. 
*idn? 
This should return the Manufacturer, Model 
number, Serial number, and Firmware version in 
the following format. 
GW, GDS-2074A, PXXXXXX, V1.00 
Note 
For further details about remote control and remote 
commands, please see the GDS-2000A programming 
manual, available on the GW Instek website.