8.1     Technical Specifications 
Specifications Assumptions 
  One year calibration cycle. 
  Operating temperature between 18°C to 28°C 
  Accuracy is expressed as: ±(% of reading + % of range ) after a 30 minute warm up and valid 
for 10 PLC (slow) 
  Temperature Coefficient: add ± [0.1% × (the applicable accuracy)/ °C] for 0°C to 18°C and 28°C 
to 40°C;   
  Relative Humidity: Up to 80% for 0°C  to  28°C  (75% RH for 10  M
Ω  and above ranges of 
resistance measurement). Up to 70 for 28°C to 40°C. 
 
Displayed Readings and Reading rates 
 
Full Scale displayed reading 
Slow 
Med 
Fast 
119,999 
119,999 
11,999 
 
Reading rates on front panel (Reading/sec, Approximate) 
Measurement Function 
Slow 
Med 
Fast 
DCV 
16 
57 
DCA 
16 
57 
ACV 
25 
ACA 
25 
2-
Wire Ω (Ranges below 120 kΩ ) 
16 
57 
2-Wi
re Ω (120 kΩ range and above) 
16 
25 
4-Wi
re Ω (Ranges below 120 kΩ) 
10 
33 
4-Wi
re Ω (120 kΩ range and above) 
10 
20 
Freq 
———— 
———— 
Period 
———— 
———— 
Diode 
———— 
16 
———— 
Continuity 
———— 
———— 
57