Basic Measurements 
22 
 
 
Chapter 3    Basic Measurements 
3.1    Overview 
The front panel has two rows of keys to select various functions and operations. Most keys have a 
shifted function printed in blue  above the key. To perform a shifted function, press 
  (the  Shift 
annunciator will turn on). Then, press the key that has the desired label above it. For example, to select 
the AC current function, press 
  then press 
  (AC I). 
 
If you accidentally press 
, just press it again to turn off the Shift annunciator. 
3.2    Measuring Voltage   
Voltage ranges: 120 mV, 1.2 V, 12 V, 120 V, 1000 V (750 VAC) 
Maximum resolution: 1 
μV (on 120 mV range) 
3.2.1    Connections 
Assuming the multimeter is under factory default conditions, the basic procedure for measuring voltage 
is as follows: 
1.  Connect test leads to INPUT HI and LO terminals. 
2.  Select DC or AC voltage measurement by pressing 
  or 
  respectively. 
3.  Press 
  to  toggle between  auto  and manual ranging. Notice the AUTO  annunciator is 
displayed with auto ranging.    For manual range, use the RANGE     
and         keys to select 
the appropriate range for measurement.   
4.  Connect test leads to the sources as shown in Figure 3-1. 
 
CAUTION:   Do not apply more than 1000 V peak
 
to the input or it will damage the instrument.   
 
5.  If the “OVR.FLW”  message is displayed, press the up        key  to  select  a higher range until a 
normal reading is displayed (or press 
  key for auto ranging). Use the lowest possible range for 
the best resolution.    The measured reading is displayed.