Basic Measurements 
29 
 
T2AL 250V
SENSE
Ω
4W
VΩ
LO
HI
10A
CATⅡ(300V)
CATⅠ(1000V)
350V
MAX
1000V
MAX
12A
MAX
mA
12A
INPUT
AC Voltage
Source
DC V
AC V
2W
Fr eq
Aut o
Tr i g
MX+B
Shi f t
Cont
Rel
¦ ¸
Period
dB/m
Fast
Menu
Recall
Med
Slow
Hold
CHOICES
LEVEL
ENTER
ESC
LOCAL
%
I
DC
I
AC
4W
Ω
FAST
MED
SLOW
ADRS
RMT
HOLD TRIG
*
MEM
AUTO
REL
FILT MATH
SHIFT
4W
ERR
POWER
R
5
5492B
1 2
/
Digit Multimeter
 
                  Input Impedance = 1 
MΩ in parallel with <100 pF 
                   
CAUTION:  Maximum Input = 750 RMS, or 1000 V Peak 
 
Figure 3-4 
Frequency and Period Measurements 
3.6    Measuring Continuity 
The multimeter uses the 1 
kΩ range to measure circuit continuity. A threshold resistance level (1 Ω to 
1000 
Ω) should be set.  The factory default value is 10 Ω. The multimeter alerts you with a beep when a 
reading is below the set level. 
 
Note: 
Continuity function defaults to FAST (0.1 PLC) rate and cannot be changed. 
3.6.1    Connections 
Assuming the multimeter is under factory default conditions, the basic procedure for continuity testing is 
as follows:   
1.  Connect test leads to the INPUT HI and LO terminals. 
2.  Select Continuity measurement function by pressing 
3.  Connect test leads to the resistance under test as shown in Figure 3-5. 
4.  The measured reading is displayed.