Basic Measurements 
35 
 
changing the reference value. 
5.  Press 
  →          key to exit the menu and return to the dB math operation. 
 
Notes:  The dB calculation takes the absolute value of the ratio V
IN
/V
REF
.  The largest negative 
value of dB is -160 dB. This will accommodate a ratio of V
IN
 = 1 µV, V
REF
 = 1000 V. 
 
 
3.8.4    dBm Calculation 
dBm is defined as decibels above or below a 1  mW reference. With a user-programmable reference 
impedance, B&K 5492B reads 0 dBm when the voltage needed to dissipate 1mW through the reference 
impedance is applied. The relationship between dBm, reference impedance, and the voltage is defined 
by the following equation: 
                    
 
Where:   V
IN
 is the DC or AC input signal voltage value. 
Z
REF
 is the specified reference impedance. 
 
If a relative value is in effect when dBm is selected, the relative value will be converted to dBm value 
before REL is applied. If REL is applied after dBm  has been selected, dBm  calculation will have REL 
applied to it directly. 
 
To set the reference impedance, perform the following steps: 
1.  Press 
  → 
, the voltage reference value for dB math function will be displayed. 
2.  Press 
  (ENTER) to confirm the voltage reference value, now you have selected the dB math 
function. 
3.  Press 
  → 
  again,  and  the  impedance reference value for dBm math function will be 
displayed: 
REF:  0075 
Ω 
4.  Use         and         keys to select cursor position and use          and   
    keys to increment 
or decrement the digits respectively.    Enter a value from 1 
Ω to 9999 Ω. 
5.  Press 
(ENTER)  to confirm the reference impedance,  and  the message “CHANGE SAVED” 
will be displayed for a moment. The multimeter will then return back to the measurement status. 
6.  Now the meter will display the result of the dBm calculation. 
 
If you want to change the impedance reference value after the dBm function is enabled, you can also do 
the following: 
1.  Press 
  →        to enter the command level, “5: dBm REF” will be displayed. 
2.  Press         to enter the parameter level, and the current  impedance  reference value will be 
displayed: 
REF: +1.0000    
3.  Use         and         keys to select cursor position and use          and   
    keys to increment 
or decrement the digits respectively.    Enter a value from 1 
Ω to 9999 Ω.