Bk Precision BK-5492B Digital-Multimeter, DMM, BK-5492B Hoja De Datos

Los códigos de productos
BK-5492B
Descargar
Página de 111
SCPI Command Reference 
92 
 
6.3.5    SYSTem Subsystem 
The SYSTem subsystem contains miscellaneous commands that are summarized in Table 6-5 
Table 6-5 
SYSTem Command Summary
 
Command 
Function Description 
Default   
:SYSTem 
  :PRESET 
  :AZERo 
    :STATe <b> 
    :STATe? 
  :BEEPer 
    [:STATe] <b> 
    [:STATe]? 
  :LOCal 
 
Return to system defaults 
Path to set up autozero 
Enable or disable autozero 
Query autozero 
Path to control beeper 
Enable or disable beeper 
Query state of beeper 
Take multimeter out of remote state and restore front panel 
operation 
 
 
 
ON 
 
 
ON 
 
 
 
:PRESET 
Command Syntax: 
:SYSTem:PRESET 
 
Description: 
This command returns the instrument to factory defaults states. 
 
:AZERo   
:STATe <b> 
Command Syntax: 
:SYSTem:AZERo:STATe <b> 
 
Command parameter: 
<b> =    1 or ON   
 
 
 
Enable autozero 
 
 
 
0 or OFF 
 
 
 
Disable autozero 
 
Query: 
:STATe?                  
 
Query state of autozero. 
Return String: 
 
1 or 0 
 
Description: 
This command is used to enable or disable autozero. When enabled, accuracy is optimized.    When 
disabled, speed is increased at the expense of accuracy. 
 
Note: Before you can enable or disable autozero, the multimeter  must first be in the idle 
state.